0.3KG Throat Swab Coronavirus Test Kit Clinical Performance
Product Details:
Place of Origin: | Beijing |
Certification: | CE; ISO |
Payment & Shipping Terms:
Minimum Order Quantity: | 10-100pcs |
---|---|
Price: | negotiable |
Packaging Details: | 25pcs per case |
Delivery Time: | 3-5working days |
Supply Ability: | 1million per week |
Detail Information |
|||
Specificity: | 100% | Overall Agreement: | 97.5% |
---|---|---|---|
Sample Type: | Nasal/Throat Swab | Assay Incubation: | 15minutes, RT |
Format: | 25T/Kit | N.W.: | 0.3KG |
Number Of Box: | 50Kits | ||
High Light: | 25T/Kit Coronavirus Test Kit,0.3KG Coronavirus Test Kit,Throat Swab Coronavirus Test Kit |
Product Description
0.3KG Throat Swab Coronavirus Test Kit Clinical Performance
COVID-19 Antigen Rapid Test Kit Clinical Performance of Antigen Test for fresh swabs
PERFORMANCE CHARACTERCS
The assay demonstrated excellent clinical specificity 100% (95% CI: 98.12%-100%). Overall Agreement was 96.21% (93.34%-97.87%).
1. Clinical Performance of Antigen Test for fresh swabs
Using 127 fresh NP swabs which were collected freshly in Extraction Buffer in the kit to evaluate the Clinical performance of the COVID-19 Antigen Rapid Test Kit. 67 Patients who presented within 7 days of symptom onset and 60 normal persons were included in the initial primary evaluation. The Result is below:
Sansure Covid-19 RT-PCR Assay | ||||
Positive | Negative | Total | ||
Kewei COVID-19 Antigen Rapid Assay | Positive | 64 | 0 | 64 |
Negative | 3 | 60 | 63 | |
Total | 67 | 60 | 127 |
The assay demonstrated acceptable clinical sensitivity for fresh swab samples is 95.52% (95% CI: 87.64%-98.47%) when compared to a molecular device produced by Sansure Inc. The assay demonstrated excellent clinical specificity 100% (95% CI: 93.98%-100%). Overall Agreement was 97.64% (93.28%-99.19%).
2. Overall Clinical Performance of Antigen Test Kit
A total of above 157 positive samples and 260 negative samples were detected to evaluate the clinical performance of the COVID-19 Antigen Rapid Test Kit. The Result is below:
Sansure Covid-19 RT-PCR Assay | ||||
Positive | Negative | Total | ||
Kewei COVID-19 Antigen Rapid Assay | Positive | 143 | 0 | 143 |
Negative | 14 | 260 | 274 | |
Total | 157 | 260 | 417 |
The assay demonstrated acceptable total clinical sensitivity is 91.08% (95% CI: 85.59%-94.61%) when compared to a molecular device produced by Sansure Inc. The assay demonstrated excellent clinical specificity 100% (95% CI: 98.54%-100%). Overall Agreement was 96.64% (94.44%-97.99%).
4. Limit of Detection(LoD)
Limit of detection (LoD) was determined by evaluating different concentrations of heat inactivated SARS-CoV-2 virus. LoD of the SARS-CoV-2 antigen rapid kit was confirmed as 38.5 TCID50/ml.
5. High-dose Hook effect
No high dose hook effect was observed when test with up to a concentration of 6.3×105 TCID50/ml of heat inactive SARS-CoV-2 virus from Academy of Military Sciences PLA China.
6. Cross Reactivity
Many types of pneumonia are companied by fever, cough and other respiratory symptoms. In order to eliminate similar clinical symptoms of other types of pneumonia effects, pneumonia mycoplasma, influenza A, pneumococcus etc. were detected for the specific assessment. Each of the organism and viruses were tested in the absence or presence of heat inactivated SARS-CoV-2 virus(308TCID 50/ml), No cross-reactivity or interference was seen.